The N2011 Multi-Use Temperature Data Logger allows easy monitoring of temperature sensitive systems, in transit or in storage. Complying with EN12830:1999 standard, it is perfectly suited for cold chain applications.
The N2011 Multi-Use Temperature Data Logger allows easy monitoring of temperature sensitive systems, in transit or in storage. Complying with EN12830:1999 standard, it is perfectly suited for cold chain applications.
With a two year warranty and a long battery life of up to 5 years, the N2011 Data Logger is designed to give long-term, reliable performance.
One of our range of Diligence EV Data Loggers, the N2011 Data Logger is also ideal for monitoring building services, scientific experiments, production processes, and heating, ventilation and air conditioning systems, and comes with a Biocote Antimicrobial Protection.
The EVTCRU Computer Docking Station (for EVT Series) is designed to work with the EVT Series Data Loggers for syncing data between the device and your PC.
The EVTCRU Computer Docking Station (for EVT Series) is designed to work with the EVT Series Data Loggers for syncing data between the device and your PC.
Name | Value |
---|---|
Sensor Type | Thermistor |
Temperature Measurement Range | -40°C to +70°C / -40°F to +158°F |
Scales | °C and °F |
Display Resolution | 0.1° |
System Accuracy °C | -25°C to +50°C ±0.5°C, -40°C to +70°C ±1°C |
System Accuracy °F | -13°F to +122°F ±1°F, -40°F to +158°F ±2°F |
Memory | 32K 16000 samples |
Communications | Via infra red interface |
Download Time | 3 minutes for 10, 000 readings (typical) |
Logging Frequency | Programmable between 1 second and 99 hours |
Internal Sensor Response Time | T90 = 15 to 30 minutes in ambient air |
Ambient Storage | -40°C to +70°C / -40°F to +158°F |
Battery Type | 1 x AA size 3.6V replaceable lithium battery |
Compatibility | Software Windows™ based software for Windows 7 |
Battery Life | Up to 5 years |
Case Material | ABS |
Environmental Protection | IP67 BS EN 60529 IEC 529 |
LED Indication | Red = Alarm, Green = Logger Active |
Dimensions | Length (including fixing lug) 88mm / 3.5ins, Width 80mm / 3.1ins, Depth 35mm / 1.4ins |